Test methods 2026_Pressalit - Flipbook - Page 2
Test methods
At Pressalit, we test our products to ensure that they can
withstand even extreme real-world conditions. We push each
product to its limits—and therefore to its weakest point—by
subjecting it to a series of rigorous tests. The products are
tested in accordance with the requirements of relevant international standards for assistive products, including:
»
»
ISO 17966 – Assistive products for personal hygiene that
support users
ISO 21856 – Assistive products
In addition, Pressalit has developed a number of its own tests
that we consider relevant to the individual products. For the
majority of Pressalit products, the ISO 17966 standard is
applied, as it sets more stringent and specific testing requirements than ISO 21856.
Weight specifications
According to the international standard for assistive products
(ISO 17966), the manufacturer must always state the maximum load of a product.
This may be specified as:
»
Maximum User Mass (MUM), kg
»
Safe Working Load (SWL), kg
On Pressalit products, weight specifications are illustrated by
icons on the product, for example on a changing table:
MUM
SWL
MUM is the abbreviation for Maximum User Mass.
SWL is the abbreviation for Safe Working Load – i.e. the
user’s weight plus, for example, an object or water that adds
to the load on the product in addition to the user.
At Pressalit, we aim to specify MUM and SWL within a range
that reflects real-world conditions in terms of realistic user
weights and wall strength. Our highest priority is to create safe
products so that users can feel secure. We take this responsibility very seriously and have therefore, among other things,
developed a guide with recommendations for installation on
different wall types.
See the fixing guide available for download on the relevant
product pages.
The ISO 17966 and ISO 21856 standards require, as a
minimum:
»
»
Strength testing (static testing)
Durability testing (dynamic testing)
The load that a wall-mounted product can withstand increases
the closer the load is applied to the wall. ISO 17966 requires
testing at the most critical points, which are often the points
furthest from the wall.
All tests on wall-mounted products are carried out on a stable
wall. The strength of the products themselves does not
depend on the type of wall, but it is important to note that the
durability and strength of the complete installation will naturally
be affected by the construction and condition of the wall. We
therefore always recommend reinforcement if the wall is not
sufficiently stable.